I tested Cervicam with real healthcare workers. Here’s the usability score.

A gloved hand writing on a clipboard, representing the usability survey.
Photo: RDNE Stock project, Pexels

Cervicam Thesis Series

Part 5 of 6

  1. 1 The most important number in childbirth is basically a guess
  2. 2 Ultrasound fixes the accuracy problem. It also costs too much.
  3. 3 Building a cervix-measuring camera probe on a Raspberry Pi budget
  4. 4 Everything that broke while building Cervicam
  5. 5 I tested Cervicam with real healthcare workers. Here’s the usability score.
  6. 6 What a proof-of-concept device can and can’t tell you

Post 3 and post 4 were about building Cervicam. This post is about handing it to people who actually do this exam for a living and seeing what they made of it.

One thing I want to be upfront about: this is a proof-of-concept, tested only on a 3D-printed cervix model, never a real patient. Even several iterations further, the idea of this being used on someone in labor is a scary proposition.

Who actually tested it

Thirteen healthcare workers at a government hospital in the Philippines, all of whom had performed at least one digital vaginal exam before, tried Cervicam and then filled out a pen-and-paper System Usability Scale (SUS) survey: the standard 10-question instrument for “does this thing actually work for the person using it.” Four were doctors, four were nurses, five were medical students; eleven were women and two were men; experience ranged evenly from less than a year to five-plus years. For what it’s worth, the instrument itself tested reliable in my analysis: a Cronbach’s alpha of 0.86, comfortably in “very reliable” territory for a survey like this.

The score, and the asterisk on it

Averaged across all thirteen respondents, Cervicam’s SUS score came out to 66.15. The conventional read on SUS scores treats 68 and above as “Above Average” and anything below that as “Below Average,”1 which puts the raw number just under the line.

That raw average is dragged down by exactly one response: a score of 32.50, low enough to count as a statistical outlier. Remove it using the standard interquartile-range treatment and the mean becomes 68.96, just over the “above average” threshold. Whether Cervicam reads as “above average” or “below average” usability genuinely depends on whether that one response stays in the average or not, which is a strange place for a device’s usability score to live.

Two box plots comparing Cervicam's mean SUS score with outliers included (66.15, labeled Below Average) versus excluded (68.96, labeled Above Average).
Figure 1. Cervicam’s mean SUS score, with and without the one outlier removed.

Who liked it, who didn’t

Broken down by occupation, the scores spread out a lot: medical students averaged 80.00, doctors 60.63, and nurses 54.38. The nurses’ number is almost entirely that one outlier’s doing. Of the four nurses who responded, three averaged 61.67, close to the doctors; the fourth, the most experienced of the group, gave it 32.50, and that single response is what pulls the whole occupation’s average down near the bottom.

There’s a pattern across the dataset beyond just that one response: scores tend to go down as years of experience go up, with medical students as the one exception (most of whom had less than a year of experience themselves). The two male respondents, a medical student and a doctor, both scored above average (82.50 and 70.00), broadly in line with the rest of the pattern rather than against it.

I presented this same research at a healthcare-education conference later that year, and the read I gave it there still holds up: medical students scoring highest across the board may say more about comfort with new technology among younger users than about the device itself, and the lower confidence scores from more experienced staff likely trace back to the probe’s rigid, early-prototype build rather than the idea behind it.

Where that actually lands it

There’s a published comparison chart that places a wide range of everyday devices and interfaces on the SUS scale: Google Search and a landline phone sit near the top, Excel near the bottom, and a cluster of medical devices (a blood glucose monitor, a syringe, a pregnancy test kit) sits in the middle of the pack.2 Even using Cervicam’s lower, outlier-included score of 66.15, it lands above the EpiPen on that same chart.

A horizontal bar chart comparing SUS scores of well-known devices and interfaces, with Cervicam highlighted between a blood glucose monitor and an EpiPen.
Figure 2. Cervicam’s SUS score against other well-known devices and interfaces.2

I’d like to say that settles it, but “better than an EpiPen” is a lower bar than it sounds, and a usability score was never going to answer the harder question anyway: whether a device like this should exist at all. That’s where the next post actually goes.

This post is adapted from my unpublished Master’s thesis, “Touch Me Not”: A Cervix Uteri’s Plea for a More Objective Cervical Dilatation Measurement — Proof of Concept for a Camera-Based Cervimeter, completed July 2024 for my MSIT at MSU-IIT. Publishing it wasn’t a requirement for the degree, so parts of it are appearing publicly here for the first time. It hasn’t been peer-reviewed, and the views and conclusions are my own. Shared under CC BY-NC, with attribution appreciated.

References

Both of these come from outside my own build/testing; everything else in this post is my own survey data, reported directly.

1 Brooke, J. (1995). SUS: A quick and dirty usability scale. Usability Evaluation in Industry, 189.

2 Betteridge, K. What Every UXC Client Should Know About SUS Scores. Bentley University User Experience Center. https://bentley.edu/centers/user-experience-center/what-every-client-should-know-about-sus-scores

Enjoyed this one?

Let J know with some claps

Leave a comment

Your email stays private. Comments are held for moderation, so yours will appear once approved.

Back to The J Files